More reliable: Fully automatic calibration procedure Compatible with (0.1、1.0 、5.0 、10.0)cm-1 conductivity cell constant, covering all types of flow between purified water and concentrated water Optional for conductivity(μS/cm)or TDS(ppm) Isolated, transferable 4-20mA current loop, any two transmission range can be set in full-scale range Instru.

More details

More info

PONPE 595 CONDUCTIVITY TRANSMITTING CONTROLLER

Application:

Widely used for pharmacy, food, beverage, paper making, environment protection,or chemical industries.

Feature : 

  • - More reliable: Fully automatic calibration procedure.
  • - Seamless automatic switch for all ranges, digital temperature compensation.
  • - Compatible with (0.1、1.0 、5.0 、10.0)cm-1 conductivity cell constant, covering all types of flow between purified water and concentrated water
  • - Optional for conductivity(μS/cm)or TDS(ppm) 
  • - Able to check instant water temperature (℃) / trans. current (4~20mA), in favor of problem solution.
  • - Isolated, transferable 4-20mA current loop, any two transmission range can be set in full-scale range.
  • - Instru./ Trans mode, Instru. mode support all 4-20mA signal regular. Trans. mode support all two wire PLC system.
  • - Flexible control application: conductivity high & low limit control/ high level exceed control and hysteresis control.
  • - LCD display provides clear technical implication of various parameter settingand status, more convenient and clear than the code prompt surface.
  • - Assisted identification provides clear technical identification for parameter setting and status.
  • - Terminals on industrialization standard, EMC design (electromagnetic compatibility), platformization in R&D, serialization in product grouping.
  • - SMT (Surface Mount Technology), AOI (Automated Optical Inspection), ICT whole board analyze, computerized FCT inspection, strict quality management and control means.
  • - All products have been tested by aging experience, A level in quality.
  • - Matched calibration, complete packaging and storage of electric meter, conductivity cell; Purchase in accordance with international ordering process.

 Specification

Model PONPE 595-L PONPE 595-H
Conductivity cell model CON595L CON595H
Power supply AC 220±10%
Conductivity Cell Constant 10.00(cm-1)
Measurement Range Conductivity (0.5~2,000) uS/cm (0.5~20.000) mS/cm, (0.5~20,000) uS/cm
TDS 0.5ppm~1,000ppm 250ppm~10,000ppm
Temp. 0~50℃ (Temp.compensation component: NTC10K)
Resolution Conductivity 1μS /cm; 0.01mS/cm
TDS 0.01ppm
Temp. 0.1℃
Accuracy Conductivity 1.5% (FS)
TDS 1.5% (FS)
Temp. ±0.5℃
Temp.compensation range 0~50℃ (with25℃ as standard)
(4∼20) mA Channels Single channel, isolated
Characteristics Reversible, fully adjustable, instrument / transmitter mode for selection
Loop resistance 400Ω(max), DC24V;
Accuracy ±0.1mA
Control contacts Channels Single channel
Relay contacts SPDT
Load capacity AC 220V/3A Max; AC 110V/5A Max; DC 24V/3A Max
Work Environment Temp: 0~50℃,  relative humidity: ≤85%RH, (none condensation)
Storage Environment Temp: 0~50℃,  relative humidity: ≤85%RH, (none condensation)
Dimensions 96 mm×96mm×105 mm (H×W×D)
Hole size 91mm×91mm (H×W)
Installation Panel mounted,fast installation